How to Integrate Our Medical Flow Measurement Solution Into Your System

1. Part one of our flow measurement solution: SonoTT™ SkyLark Series

SkyLark

  • Universally integrable evaluation electronic according to standard IEC 60601
  • Class C software built to IEC 62304 specs
  • Suitable for use with class III medical devices
  • Small frame size
  • MDR-ready

 

2.Part two of our flow measurement solution: SonoTT™ Clamp-On SL

Clamp-On

  • Non-invasive operative sensor based on acoustic measurement method
  • To be used in combination with the board for an integrated blood flow measurement in existing extracorporeal tubing systems
  • Easy and hygienic attachment to all commonly used medical tubing sizes and materials
  • Waterproof: easy to clean and disinfect for long-term use

 

3. Customization Process


Benefit from an established process to customize our flow components due to your requirements:

  • Step 1: Request and requirements
  • Step 2: Samples and prototyping
  • Step 3: Onsite testing and adjustments
  • Step 4: Final changes and pilot series
  • Step 5: Serial production

4. Integration


Keep the following key aspects in mind when integrating our solutions into your application:

  • Housing: use a metal case and metal screws
  • Wiring: do not cross the cables or run them directly over or under the board
  • Power supply: use stable 3.3V, provided via three individual pins with min 300mA
  • Zero flow adjustment: carry out before each new measurement